Venturi Meter Flow Calculator
Flow through a venturi meter from the pressure drop between the inlet and the throat. Uses the ISO 5167 discharge equation with a coefficient you can adjust.
Enter inlet and throat diameters, the measured differential pressure and the fluid density. All diameters are internal.
How it works
A venturi meter narrows the flow from the inlet diameter D₁ to a throat D₂. The pressure drop ΔP between the two sections is converted to flow with the ISO 5167 equation:
Q = Cₕ · (π/4) D₂² / √(1 − β⁴) · √(2ΔP/ρ), with β = D₂/D₁.
The discharge coefficient Cₕ is about 0.984 for a machined convergent venturi and about 0.995 for a rough-cast one. The equation assumes steady, incompressible flow and a horizontal meter (elevation head is already included in the measured ΔP when tappings are at the same height).
Worked example
D₁ = 100 mm, D₂ = 50 mm, ΔP = 20 kPa, water at 1000 kg/m³, Cₕ = 0.984. β = 0.5, throat area = 0.0019635 m². Q = 0.984 × 0.0019635 / √(1 − 0.0625) × √(2 × 20000/1000) = 0.01262 m³/s = 45.4 m³/h. Throat velocity = 6.43 m/s.
Assumptions and limitations
- Incompressible flow; for gases with ΔP/p₁ above about 2% an expansibility factor is needed.
- Fully developed turbulent flow (throat Reynolds number above roughly 2 × 10⁵).
- Pressure tappings and installation per ISO 5167; poor upstream straight lengths raise the uncertainty.